Monday, January 6, 2014

#166 - PPM O2 – Part 2

We talked a little last time about oxygen analysis at the parts-per-million (ppm) level. Electrochemical sensors perform well in many applications at those levels.

Leaks and diffusion create challenges with ppm O2 applications. In a neutral pressure environment, the gas analyzer will require a built-in pump. Activation of the pump will create a slightly lower pressure in the sample tubing ahead of the pump and will promote inbound leaks at imperfect tubing connections.

Gas diffusion may also occur irrespective of pressure through polymer-type tubing based on the size of the gas molecule and the structure of the polymer. Flexible polymer tubing may initially show decent diffusion resistance. However, other constituents in the sample may alter the chemistry and structure of the tubing material and make it more prone to gas diffusion over time.

In many gas analysis applications small leaks and gas diffusion may not cause significant analytical problems. But with ppmO2 applications, these problems can have a noticeable effect. Even a small inbound leak of ambient air will introduce a high level of non-sample O2 and result in falsely high readings. It may also prevent an accurate Zero calibration. O2 diffusion may cause the same effects as inbound leaks, but at a slower rate.

The most effective solutions to these issues involve removing leaks and points of diffusion. PPM O2 analyzers should be meticulously built using stainless steel tubing with reliable connections. We prefer that ppmO2 applications have a positive sample pressure. This removes low pressure points before the sensor and allows us to use a regulator in place of a pump. Simple and strong seem to be the best methods for ppmO2.


We are currently building a couple instruments with dual-channel
PPM / % O2 analysis. The ppm channel can be switched to protect
the sensor automatically or manually depending on the customer’s
preference. In this photo, the instrument is being purged with N2
in advance of factory calibration.

The ppm side is being purged out with N2
in preparation for factory calibration.
Notice that the tubing is narrow diameter
flexible stainless steel to minimize leaks
or diffusion during this process.

In most applications, a ppm analysis requirement implies a high proportion of other gases in the sample. Some gases interfere with accurate ppm analysis of other gases. So it is important that we have a good understanding of the sample gas make-up before designing an analyzer.

We have produced portable and a continuous PPM O2 analyzers which may be suitable for your application.

1-800-295-3771
sales at nova-gas dot com
websales at nova-gas dot com
http://www.nova-gas.com/

Thursday, December 12, 2013

#164 - PPM O2 Analysis – Part 1

We have had quite a few inquiries lately about oxygen analysis in the parts-per-million (PPM) scale.

The ambient air we breathe has 20.95% oxygen (O2) in it. This proportion of O2 is suitable for sustaining many forms of life and biological processes on this planet. However, some processes require atmospheres of low O2 or even trace O2. When verifying the purity of a produced gas, instead of measuring the target gas, it is sometimes best to measure the impurities. If the impurity is known to be O2, measuring the O2 will make good sense. Instead of being stated as a percent (%) O2, some very low O2 atmospheres are stated in parts-per-million (PPM).

Measuring ppmO2 can be a challenge for several reasons:
  • PPM measurements require a suitable degree of accuracy and
    precision to be meaningful.
  • Not all O2 measurement technologies are sensitive enough to
    function at low ppmO2 levels.
  • Ambient O2 is ubiquitous and is difficult to exclude.

In our experience, the electrochemical sensor is most commonly requested by our customers for ppmO2 measurement. In Nova’s product line, this methodology can be used for any range from 0-100 PPM to 0-19,999 PPM O2. The electrochemical sensor provides good sensitivity, accuracy, and speed of response to O2 levels in these ranges.

One disadvantage of electrochemical ppmO2 sensors is their sensitivity to high levels of O2. If a sensor is set up for a range of 0-1,000ppm and it is exposed to ambient air, it will be ‘blinded’ temporarily. The effect is similar to stepping out into bright sunlight after spending time in a low-light environment. An electrochemical sensor may have long recovery time after the exposure for a couple of reasons:
  • Chemical recovery of the sensor and removal of high O2
    from the cell’s diffusion layer.
  • Purging the excess O2 from the internal sample tubing
    and related components.

Therefore, it is important to avoid exposing the sensor to ambient air. This may not be too difficult with a continuous analyzer that is permanently connected to a low O2 environment. However, a portable analyzer may not have this advantage.

In these analyzers, we internally protect the O2 sensors by purging out the ambient air out before exposing the sensors to the sample. When finished analyzing, we trap the last ppmO2 sample in the sensor before decoupling the analyzer from the process. We do both of these operations by using PURGE / SAMPLE switch on the front panel.

Before the analyzer is exposed to the sample gas,
the switch on analyzer is moved to the PURGE position.
Running the analyzer in this mode allows some time
for the ambient air to purge out of the sample tubing.
After the purge time, the switch can be moved over to SAMPLE.

The Model 325 is designed for ppmO2 analysis. If a system like this is of interest to you, contact Nova for details.