In terms of customer interaction with Nova
continuous analyzers, there are surprisingly few requirements. Especially on
auto-cal equipped units, the instrument essentially calibrates and runs itself.
During installation and set-up, there are a
few connections for the customer to be aware of. We can roughly categorize them
as Power, Signal, and Gas.
Power
The vast majority of our continuous
analyzers are AC powered using 115VAC / 60Hz or 220VAC / 50Hz. This is dictated
by the country / region the instrument will be installed into. AC power is
brought into the analyzer cabinet and terminated at the main terminal block
located on the upper right hand side.
In some cases, a separate power supply may
be brought into the cabinet if the customer has required an internal cabinet
LEL monitor that can depower the analyzer during an alarm condition. The
separate power supply allows the LEL monitor to continue functioning when the
rest of the analyzer is depowered.
Signal
The analyzer will have a visual display of
the gas readings. The most common format of communicating the gas readings to
other parts of the customer’s instrumentation system is via 4-20mA output. The
gas analysis range is co-related to a 4-20mA scale. There will be a separate
output for each gas analyzed. The outputs are usually connected to the
customer’s PLC or data collection system. The main terminal block will have the
output connections clearly labeled as shown in the diagram.
Digital outputs such as RS-485, RS-232, and
USB are also available on some Nova analyzers. The connection points will be on
the main terminal block next to the 4-20mA outputs. We can also offer MODBUS
protocol over Ethernet connection using a small convertor usually located under
the main terminal block.
Gas
Obviously, there will be tubing connections
on the analyzer to allow flow of sample gas or calibration gas.
The SAMPLE IN port is used during normal
operation when the instrument is continuously measuring the gas. The tubing is
connected to this port directly from the process or from another sampling
component located between the process and analyzer. The SPAN and ZERO ports are
used during calibration. The DRAIN port is located on the bottom of some
analyzers and is used to evacuate condensate from the analyzer.
The tubing attached to the SAMPLE IN and
DRAIN ports will usually contain various amounts of condensed water which can
freeze if the analyzer is installed outside in cold weather. These tubing lines
require heat-tracing to prevent freezing. (Note: the tubing shown in the picture is just temporary flexible tubing used during the manufacturing stages. The tubing used in on-site permanent installation will be stainless steel, PFTE, or another specified material.)
For information on these and other gas
analyzer systems, give Mike or Dave at Nova a call, or send us an e-mail.
1-800-295-3771
sales at nova-gas dot comwebsales at nova-gas dot com
http://www.nova-gas.com/
https://twitter.com/NOVAGAS
http://www.linkedin.com/company/nova-analytical-systems-inc-
http://www.tenovagroup.com/
-
No comments:
Post a Comment